Goal Reached Thanks to every supporter — we hit 100%!

Goal: 1000 CNY · Raised: 1000 CNY

100.0%

CVE-2024-9834— Improper data protection on Life2000 ventilator serial interface

CVSS 9.3 · Critical EPSS 0.05% · P14
Get alerts for future matching vulnerabilitiesLog in to subscribe

I. Basic Information for CVE-2024-9834

Vulnerability Information

Have questions about the vulnerability? See if Shenlong's analysis helps!
View Shenlong Deep Dive ↗

Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.

Vulnerability Title
Improper data protection on Life2000 ventilator serial interface
Source: NVD (National Vulnerability Database)
Vulnerability Description
Improper data protection on the ventilator's serial interface could allow an attacker to send and receive messages that result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
Source: NVD (National Vulnerability Database)
CVSS Information
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Source: NVD (National Vulnerability Database)
Vulnerability Type
敏感数据的明文传输
Source: NVD (National Vulnerability Database)
Vulnerability Title
Baxter Life2000 安全漏洞
Source: CNNVD (China National Vulnerability Database)
Vulnerability Description
Baxter Life2000是Baxter公司的一款无需面罩的无创呼吸机。 Baxter Life2000存在安全漏洞,该漏洞源于呼吸机串行接口上的数据保护不当,可能会允许攻击者发送和接收对设备设置和性能产生意外影响的消息,导致未经授权的信息泄露。
Source: CNNVD (China National Vulnerability Database)
CVSS Information
N/A
Source: CNNVD (China National Vulnerability Database)
Vulnerability Type
N/A
Source: CNNVD (China National Vulnerability Database)

Shenlong Deep Dive — AI Deep Analysis

10-question deep dive: root cause, exploitation, mitigation, urgency. Read summary free, full version requires login.

Affected Products

VendorProductAffected VersionsCPESubscribe
BaxterLife2000 Ventilation System 06.08.00.00 and prior -

II. Public POCs for CVE-2024-9834

#POC DescriptionSource LinkShenlong Link
AI-Generated POCPremium

No public POC found.

Login to generate AI POC

III. Intelligence Information for CVE-2024-9834

登录查看更多情报信息。

Same Patch Batch · Baxter · 2024-11-14 · 8 CVEs total

CVE-2024-4896610.0 CRITICALLife2000 service tools for test and calibration do not support user authentication
CVE-2024-4896710.0 CRITICALLife2000 ventilator and Service PC lack sufficient audit logging capabilities
CVE-2024-98329.3 CRITICALNo limit on failed login attempts with Clinician Password or Serial Number Clinician Passw
CVE-2024-489749.3 CRITICALLife2000 Ventilator does not perform proper file integrity checks when adopting firmware u
CVE-2024-489719.3 CRITICALClinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventila
CVE-2024-489709.3 CRITICALLife2000 Ventilator microcontroller lacks memory protection
CVE-2024-489739.3 CRITICALDebug port on Life2000 Ventilator serial interface is enabled by default

IV. Related Vulnerabilities

V. Comments for CVE-2024-9834

No comments yet


Leave a comment