CVE-2024-48970— Life2000 Ventilator microcontroller lacks memory protection

CVSS 9.3 · Critical EPSS 0.09% · P25
Get alerts for future matching vulnerabilitiesLog in to subscribe

I. Basic Information for CVE-2024-48970

Vulnerability Information

Have questions about the vulnerability? See if Shenlong's analysis helps!
View Shenlong Deep Dive ↗

Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.

Vulnerability Title
Life2000 Ventilator microcontroller lacks memory protection
Source: NVD (National Vulnerability Database)
Vulnerability Description
The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorized information disclosure.
Source: NVD (National Vulnerability Database)
CVSS Information
CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
Source: NVD (National Vulnerability Database)
Vulnerability Type
CWE-1191
Source: NVD (National Vulnerability Database)
Vulnerability Title
Baxter Life2000 安全漏洞
Source: CNNVD (China National Vulnerability Database)
Vulnerability Description
Baxter Life2000是Baxter公司的一款无需面罩的无创呼吸机。 Baxter Life2000 06.08.00.00及之前版本存在安全漏洞,该漏洞源于缺少内存保护,允许攻击者通过JTAG接口读取或写入闪存,可能会破坏设备的功能或导致未经授权的信息泄露。
Source: CNNVD (China National Vulnerability Database)
CVSS Information
N/A
Source: CNNVD (China National Vulnerability Database)
Vulnerability Type
N/A
Source: CNNVD (China National Vulnerability Database)

Shenlong Deep Dive — AI Deep Analysis

10-question deep dive: root cause, exploitation, mitigation, urgency. Read summary free, full version requires login.

Affected Products

VendorProductAffected VersionsCPESubscribe
BaxterLife2000 Ventilation System 06.08.00.00 and prior -

II. Public POCs for CVE-2024-48970

#POC DescriptionSource LinkShenlong Link
AI-Generated POCPremium

No public POC found.

Login to generate AI POC

III. Intelligence Information for CVE-2024-48970

登录查看更多情报信息。

Same Patch Batch · Baxter · 2024-11-14 · 8 CVEs total

CVE-2024-4896610.0 CRITICALLife2000 service tools for test and calibration do not support user authentication
CVE-2024-4896710.0 CRITICALLife2000 ventilator and Service PC lack sufficient audit logging capabilities
CVE-2024-98329.3 CRITICALNo limit on failed login attempts with Clinician Password or Serial Number Clinician Passw
CVE-2024-98349.3 CRITICALImproper data protection on Life2000 ventilator serial interface
CVE-2024-489749.3 CRITICALLife2000 Ventilator does not perform proper file integrity checks when adopting firmware u
CVE-2024-489719.3 CRITICALClinician Password and Serial Number Clinician Password are hard-coded in Life2000 Ventila
CVE-2024-489739.3 CRITICALDebug port on Life2000 Ventilator serial interface is enabled by default

IV. Related Vulnerabilities

V. Comments for CVE-2024-48970

No comments yet


Leave a comment