Goal Reached Thanks to every supporter — we hit 100%!

Goal: 1000 CNY · Raised: 1000 CNY

100.0%

CVE-2017-12725

EPSS 0.18% · P39
Get alerts for future matching vulnerabilitiesLog in to subscribe

I. Basic Information for CVE-2017-12725

Vulnerability Information

Have questions about the vulnerability? See if Shenlong's analysis helps!
View Shenlong Deep Dive ↗

Although we use advanced large model technology, its output may still contain inaccurate or outdated information.Shenlong tries to ensure data accuracy, but please verify and judge based on the actual situation.

Vulnerability Title
N/A
Source: NVD (National Vulnerability Database)
Vulnerability Description
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
Source: NVD (National Vulnerability Database)
CVSS Information
N/A
Source: NVD (National Vulnerability Database)
Vulnerability Type
N/A
Source: NVD (National Vulnerability Database)
Vulnerability Title
Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump 安全漏洞
Source: CNNVD (China National Vulnerability Database)
Vulnerability Description
Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump是英国史密斯医疗(Smiths Medical)公司的一款注射器输液泵。 Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump 1.1版本、1.5版本和1.6版本中存在安全漏洞,该漏洞源于程序使用了硬编码证书来自动创建无线网络连接。远程攻击者可利用该漏洞绕过安全限制。
Source: CNNVD (China National Vulnerability Database)
CVSS Information
N/A
Source: CNNVD (China National Vulnerability Database)
Vulnerability Type
N/A
Source: CNNVD (China National Vulnerability Database)

Affected Products

VendorProductAffected VersionsCPESubscribe
-Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump -

II. Public POCs for CVE-2017-12725

#POC DescriptionSource LinkShenlong Link
AI-Generated POCPremium

No public POC found.

Login to generate AI POC

III. Intelligence Information for CVE-2017-12725

登录查看更多情报信息。

Same Patch Batch · n/a · 2018-02-15 · 31 CVEs total

CVE-2017-18088Atlassian Bitbucket Server 输入验证漏洞
CVE-2018-0864Microsoft Project Server和SharePoint Enterprise Server 权限许可和访问控制漏洞
CVE-2018-0820Microsoft Windows kernel 权限许可和访问控制问题漏洞
CVE-2018-7057Steelcase RoomWizard 跨站脚本漏洞
CVE-2018-7056Steelcase RoomWizard 信息泄露漏洞
CVE-2018-7055Steelcase RoomWizard 安全漏洞
CVE-2018-54403S-Smart CODESYS和CODESYS runtime system 缓冲区错误漏洞
CVE-2017-18189Sound eXchange 安全漏洞
CVE-2017-12726Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump 安全漏洞
CVE-2017-12724Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump 安全漏洞
CVE-2017-12723Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump 安全漏洞
CVE-2017-12722Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump 安全漏洞
CVE-2017-12721Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump 安全漏洞
CVE-2017-12720Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump 访问控制错误漏洞
CVE-2017-12718Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump 缓冲区错误漏洞
CVE-2018-5767Tenda AC15 输入验证漏洞
CVE-2018-7169shadow 安全特征问题漏洞
CVE-2018-7054Irssi 安全漏洞
CVE-2018-7053Irssi 安全漏洞
CVE-2018-7052Irssi 安全漏洞

Showing top 20 of 31 CVEs. View all on vendor page → →